Unknown

Dataset Information

0

Can the ETView VivaSight SL Rival Conventional Intubation Using the Macintosh Laryngoscope During Adult Resuscitation by Novice Physicians?: A Randomized Crossover Manikin Study.


ABSTRACT: The aim of this study was to assess the performance of the ETView VivaSight SL (ETView) single-lumen airway tube with an integrated high-resolution imaging camera in a manikin-simulated cardiopulmonary resuscitation scenario with and without chest compression. This was a randomized crossover manikin trial. Following a brief training session, 107 volunteer novice physicians who were inexperienced with airway management attempted to intubate a manikin using a Macintosh laryngoscope (MAC) and an ETView, with and without chest compressions. The participants were instructed to make 3 attempts in each scenario. In this trial, we compared intubation time, intubation success rates, and glottic visibility using a Cormack & Lehane Grade. Dental compression and ease of use of each device were also assessed. Median intubation times for the ETView and MAC without chest compressions were 17 (IQR, 15-19) s and 27 (IQR, 25-33) s, respectively (P?

SUBMITTER: Kurowski A 

PROVIDER: S-EPMC4616410 | BioStudies | 2015-01-01

REPOSITORIES: biostudies

Similar Datasets

2015-01-01 | S-EPMC4575358 | BioStudies
2018-01-01 | S-EPMC5986950 | BioStudies
2016-01-01 | S-EPMC4873178 | BioStudies
2018-01-01 | S-EPMC6133438 | BioStudies
2016-01-01 | S-EPMC5591099 | BioStudies
2016-01-01 | S-EPMC4773040 | BioStudies
2008-01-01 | S-EPMC2614464 | BioStudies
2017-01-01 | S-EPMC5215847 | BioStudies
2017-01-01 | S-EPMC5576690 | BioStudies
2008-01-01 | S-EPMC2613429 | BioStudies