<HashMap><database>biostudies-literature</database><scores/><additional><submitter>Oertli JM</submitter><funding>Swiss National Science Foundation</funding><funding>Wellcome Trust</funding><pagination>18</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC10732091</full_dataset_link><repository>biostudies-literature</repository><omics_type>Unknown</omics_type><volume>12(12)</volume><pubmed_abstract>&lt;h4>Purpose&lt;/h4>The purpose of this study was to establish and validate a novel fundus-controlled dark-adaptometry method.&lt;h4>Methods&lt;/h4>We developed a custom dark-adaptometry software for the S-MAIA device using the open-perimetry-interface. In the validation-substudy, participants underwent dark-adaptometry testing with a comparator device (MonCvONE, 59% rhodopsin bleach, cyan and red stimuli centered at 2 degrees, 4 degrees, and 6 degrees eccentricity). Following a brief break (approximately 5 minutes), the participants were bleached again and underwent dark-adaptometry testing with the S-MAIA device (same loci). In the retest reliability-substudy, participants were tested twice with the S-MAIA device (same loci as above). Nonlinear curve fitting was applied to extract dark-adaptation </pubmed_abstract><journal>Translational vision science &amp; technology</journal><pubmed_title>Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study.</pubmed_title><pmcid>PMC10732091</pmcid><funding_grant_id>201165</funding_grant_id><funding_grant_id>310030</funding_grant_id><pubmed_authors>Oertli JM</pubmed_authors><pubmed_authors>Jeffrey BG</pubmed_authors><pubmed_authors>Pfau M</pubmed_authors><pubmed_authors>Pfau K</pubmed_authors><pubmed_authors>Scholl HPN</pubmed_authors></additional><is_claimable>false</is_claimable><name>Establishing Fully-Automated Fundus-Controlled Dark Adaptometry: A Validation and Retest-Reliability Study.</name><description>&lt;h4>Purpose&lt;/h4>The purpose of this study was to establish and validate a novel fundus-controlled dark-adaptometry method.&lt;h4>Methods&lt;/h4>We developed a custom dark-adaptometry software for the S-MAIA device using the open-perimetry-interface. In the validation-substudy, participants underwent dark-adaptometry testing with a comparator device (MonCvONE, 59% rhodopsin bleach, cyan and red stimuli centered at 2 degrees, 4 degrees, and 6 degrees eccentricity). Following a brief break (approximately 5 minutes), the participants were bleached again and underwent dark-adaptometry testing with the S-MAIA device (same loci). In the retest reliability-substudy, participants were tested twice with the S-MAIA device (same loci as above). Nonlinear curve fitting was applied to extract dark-adaptation </description><dates><release>2023-01-01T00:00:00Z</release><publication>2023 Dec</publication><modification>2026-05-28T19:11:47.962Z</modification><creation>2025-04-05T20:12:24.762Z</creation></dates><accession>S-EPMC10732091</accession><cross_references><pubmed>38112496</pubmed><doi>10.1167/tvst.12.12.18</doi></cross_references></HashMap>