{"database":"biostudies-literature","file_versions":[],"scores":null,"additional":{"omics_type":["Unknown"],"volume":["15(1)"],"submitter":["Mouawad L"],"pubmed_abstract":["A sensitive and reliable electrochemical biosensor for the detection of benzalkonium chloride (BAC) and didecyldimethylammonium chloride (DDAC), the most commonly used disinfectant biocides in the agri-food industry, is described. Acetylcholinesterase from <i>Drosophila melanogaster</i> (DM AChE) and butyrylcholinesterase from horse serum (BChE) were immobilized by entrapment in a photocrosslinkable polymer on the surface of carbon screen-printed electrodes. Preliminary tests conducted in phosphate buffer showed limits of detection (LODs) of 0.26 µM for BAC using the BChE-based sensor and 0.04 µM for DDAC using the DM AChE sensor. These performances comply with the European regulation for dairy products, which sets a maximum allowable concentration of 0.28 µM for biocides. However, when te"],"journal":["Biosensors"],"pagination":["26"],"full_dataset_link":["https://www.ebi.ac.uk/biostudies/studies/S-EPMC11763983"],"repository":["biostudies-literature"],"pubmed_title":["Enhancing Biocide Safety of Milk Using Biosensors Based on Cholinesterase Inhibition."],"pmcid":["PMC11763983"],"pubmed_authors":["Catanante G","Noguer T","Mouawad L","Istamboulie G"],"additional_accession":[]},"is_claimable":false,"name":"Enhancing Biocide Safety of Milk Using Biosensors Based on Cholinesterase Inhibition.","description":"A sensitive and reliable electrochemical biosensor for the detection of benzalkonium chloride (BAC) and didecyldimethylammonium chloride (DDAC), the most commonly used disinfectant biocides in the agri-food industry, is described. Acetylcholinesterase from <i>Drosophila melanogaster</i> (DM AChE) and butyrylcholinesterase from horse serum (BChE) were immobilized by entrapment in a photocrosslinkable polymer on the surface of carbon screen-printed electrodes. Preliminary tests conducted in phosphate buffer showed limits of detection (LODs) of 0.26 µM for BAC using the BChE-based sensor and 0.04 µM for DDAC using the DM AChE sensor. These performances comply with the European regulation for dairy products, which sets a maximum allowable concentration of 0.28 µM for biocides. However, when te","dates":{"release":"2025-01-01T00:00:00Z","publication":"2025 Jan","modification":"2025-04-04T22:07:48.535Z","creation":"2025-04-04T22:07:48.535Z"},"accession":"S-EPMC11763983","cross_references":{"pubmed":["39852077"],"doi":["10.3390/bios15010026"]}}