<HashMap><database>biostudies-literature</database><scores/><additional><submitter>Goodburn R</submitter><funding>Cancer Research UK</funding><pagination>28-40</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC12021324</full_dataset_link><repository>biostudies-literature</repository><omics_type>Unknown</omics_type><volume>94(1)</volume><pubmed_abstract>&lt;h4>Purpose&lt;/h4>To measure the gradient system transfer function (GSTF) of an MR-Linac (Elekta Unity, Stockholm, Sweden) using an accessible phantom-based method and to apply trajectory corrections for UTE image reconstruction in the context of MR-guided radiotherapy of lung cancer.&lt;h4>Methods&lt;/h4>The first-order GSTF of a 1.5 T, split gradient Elekta Unity MR-Linac was measured using a thin-slice technique to characterize gradient system imperfections for each physical gradient axis (X, Y, Z). Repeatability measurements of the GSTF were performed 48 h apart. The GSTF was applied to trajectory correction in multi-echo UTE image reconstruction (TEs = 0.176, 1.85, 3.52 ms) to allow for UTE-Dixon inputs in the generation of synthetic CT. Images were acquired in an anthropomorphic phantom and </pubmed_abstract><journal>Magnetic resonance in medicine</journal><pubmed_title>Gradient system characterization of a 1.5 T MR-Linac with application to 4D UTE imaging for adaptive MR-guided radiotherapy of lung cancer.</pubmed_title><pmcid>PMC12021324</pmcid><funding_grant_id>A26234</funding_grant_id><funding_grant_id>C33589/A28284</funding_grant_id><pubmed_authors>Barnes H</pubmed_authors><pubmed_authors>Oelfke U</pubmed_authors><pubmed_authors>Goodburn R</pubmed_authors><pubmed_authors>Bruijnen T</pubmed_authors><pubmed_authors>Wetscherek A</pubmed_authors><pubmed_authors>Nair P</pubmed_authors><pubmed_authors>Ahmed M</pubmed_authors><pubmed_authors>Lecoeur B</pubmed_authors></additional><is_claimable>false</is_claimable><name>Gradient system characterization of a 1.5 T MR-Linac with application to 4D UTE imaging for adaptive MR-guided radiotherapy of lung cancer.</name><description>&lt;h4>Purpose&lt;/h4>To measure the gradient system transfer function (GSTF) of an MR-Linac (Elekta Unity, Stockholm, Sweden) using an accessible phantom-based method and to apply trajectory corrections for UTE image reconstruction in the context of MR-guided radiotherapy of lung cancer.&lt;h4>Methods&lt;/h4>The first-order GSTF of a 1.5 T, split gradient Elekta Unity MR-Linac was measured using a thin-slice technique to characterize gradient system imperfections for each physical gradient axis (X, Y, Z). Repeatability measurements of the GSTF were performed 48 h apart. The GSTF was applied to trajectory correction in multi-echo UTE image reconstruction (TEs = 0.176, 1.85, 3.52 ms) to allow for UTE-Dixon inputs in the generation of synthetic CT. Images were acquired in an anthropomorphic phantom and </description><dates><release>2025-01-01T00:00:00Z</release><publication>2025 Jul</publication><modification>2025-07-01T03:05:32.485Z</modification><creation>2025-07-01T03:05:32.485Z</creation></dates><accession>S-EPMC12021324</accession><cross_references><pubmed>40106794</pubmed><doi>10.1002/mrm.30505</doi></cross_references></HashMap>