<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>26(9)</volume><submitter>Li B</submitter><pubmed_abstract>&lt;h4>Background&lt;/h4>Ultra-wide coverage CT (> 128 detector rows) makes it possible to image a heart or brain in a single rotation but are associated with large cone angles, which can severely degrade the image quality.&lt;h4>Purpose&lt;/h4>This study evaluate the image quality and artifact levels of a dual-focal-spot single-detector (DFSSD) CT geometry designed to achieve 140 mm z-axis coverage, through a simulation study.&lt;h4>Methods&lt;/h4>The DFSSD CT system employs two x-ray focal spots spaced 90 mm apart along the z-axis and a 100 mm CT detector. It provides 140 mm z-axis coverage at the iso-center. A research CT simulation package (CatSim) was employed throughout this study. To compare the image quality, three system geometries were simulated: (A) single focal spot with 40 mm detector (VCT40), </pubmed_abstract><journal>Journal of applied clinical medical physics</journal><pagination>e70220</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC12401943</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>Dual-focal-spot single-detector CT: A simulation study to assess cone-beam artifacts and noise.</pubmed_title><pmcid>PMC12401943</pmcid><pubmed_authors>Li B</pubmed_authors><pubmed_authors>Tang X</pubmed_authors></additional><is_claimable>false</is_claimable><name>Dual-focal-spot single-detector CT: A simulation study to assess cone-beam artifacts and noise.</name><description>&lt;h4>Background&lt;/h4>Ultra-wide coverage CT (> 128 detector rows) makes it possible to image a heart or brain in a single rotation but are associated with large cone angles, which can severely degrade the image quality.&lt;h4>Purpose&lt;/h4>This study evaluate the image quality and artifact levels of a dual-focal-spot single-detector (DFSSD) CT geometry designed to achieve 140 mm z-axis coverage, through a simulation study.&lt;h4>Methods&lt;/h4>The DFSSD CT system employs two x-ray focal spots spaced 90 mm apart along the z-axis and a 100 mm CT detector. It provides 140 mm z-axis coverage at the iso-center. A research CT simulation package (CatSim) was employed throughout this study. To compare the image quality, three system geometries were simulated: (A) single focal spot with 40 mm detector (VCT40), </description><dates><release>2025-01-01T00:00:00Z</release><publication>2025 Sep</publication><modification>2026-05-29T18:46:32.802Z</modification><creation>2026-04-08T05:43:50.52Z</creation></dates><accession>S-EPMC12401943</accession><cross_references><pubmed>40890952</pubmed><doi>10.1002/acm2.70220</doi></cross_references></HashMap>