<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>6(3)</volume><submitter>Liebanas Garcia P</submitter><pubmed_abstract>&lt;h4>Objectives&lt;/h4>The determination of urinary albumin, crucial for chronic kidney disease (CKD) diagnosis and monitoring, typically employs quantitative techniques as the gold standard. However, semi-quantitative methods such as urine reagent strips show promise as cost-effective and rapid screening tools. This study aims to evaluate the feasibility of utilizing Meditape UC-11A strips (Sysmex, Kobe, Japan) for albuminuria screening compared to quantitative assays measuring urine albumin-to-creatinine ratio (ACR).&lt;h4>Methods&lt;/h4>We systematically analyzed 1,627 strip samples using Meditape UC-11A strips whenever quantitative urinary albumin measurement was required. We compared the ACR measured by strip test with the gold standard (quantitative method determined by Atellica, Siemens, Marb</pubmed_abstract><journal>Advances in laboratory medicine</journal><pagination>345-350</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC12446912</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>Evaluation of urine albumin-to-creatinine ratio analysis using strip test as a screening method for urinary albumin determination in primary care.</pubmed_title><pmcid>PMC12446912</pmcid><pubmed_authors>Montolio Breva S</pubmed_authors><pubmed_authors>Sans Mateu T</pubmed_authors><pubmed_authors>Liebanas Garcia P</pubmed_authors></additional><is_claimable>false</is_claimable><name>Evaluation of urine albumin-to-creatinine ratio analysis using strip test as a screening method for urinary albumin determination in primary care.</name><description>&lt;h4>Objectives&lt;/h4>The determination of urinary albumin, crucial for chronic kidney disease (CKD) diagnosis and monitoring, typically employs quantitative techniques as the gold standard. However, semi-quantitative methods such as urine reagent strips show promise as cost-effective and rapid screening tools. This study aims to evaluate the feasibility of utilizing Meditape UC-11A strips (Sysmex, Kobe, Japan) for albuminuria screening compared to quantitative assays measuring urine albumin-to-creatinine ratio (ACR).&lt;h4>Methods&lt;/h4>We systematically analyzed 1,627 strip samples using Meditape UC-11A strips whenever quantitative urinary albumin measurement was required. We compared the ACR measured by strip test with the gold standard (quantitative method determined by Atellica, Siemens, Marb</description><dates><release>2025-01-01T00:00:00Z</release><publication>2025 Sep</publication><modification>2026-06-04T07:28:00.282Z</modification><creation>2026-06-04T03:05:40.233Z</creation></dates><accession>S-EPMC12446912</accession><cross_references><pubmed>40977823</pubmed><doi>10.1515/almed-2024-0068</doi></cross_references></HashMap>