<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>7(1)</volume><submitter>Lommen A</submitter><pubmed_abstract>A strategy, detailed methodology description and software are given with which the mass accuracy of U-HPLC-Orbitrap data (resolving power 50,000 FWHM) can be enhanced by an order of magnitude to sub-ppm levels. After mass accuracy enhancement all 211 reference masses have mass errors within 0.5 ppm; only 14 of these are outside the 0.2 ppm error margin. Further demonstration of mass accuracy enhancement is shown on a pre-concentrated urine sample in which evidence for 89 (342 ions) potential hydroxylated and glucuronated DHEA-metabolites is found. Although most DHEA metabolites have low-intensity mass signals, only 11 out of 342 are outside the ±1 ppm error envelop; 272 mass signals have errors below 0.5 ppm (142 below 0.2 ppm). The methodology consists of: (a) a multiple internal lock cor</pubmed_abstract><journal>Metabolomics : Official journal of the Metabolomic Society</journal><pagination>15-24</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC3040356</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>Ultra-fast searching assists in evaluating sub-ppm mass accuracy enhancement in U-HPLC/Orbitrap MS data.</pubmed_title><pmcid>PMC3040356</pmcid><pubmed_authors>Lommen A</pubmed_authors><pubmed_authors>Ruiz-Aracama A</pubmed_authors><pubmed_authors>Peters RJ</pubmed_authors><pubmed_authors>Gerssen A</pubmed_authors><pubmed_authors>Oosterink JE</pubmed_authors><pubmed_authors>Mol HG</pubmed_authors><pubmed_authors>Kools HJ</pubmed_authors></additional><is_claimable>false</is_claimable><name>Ultra-fast searching assists in evaluating sub-ppm mass accuracy enhancement in U-HPLC/Orbitrap MS data.</name><description>A strategy, detailed methodology description and software are given with which the mass accuracy of U-HPLC-Orbitrap data (resolving power 50,000 FWHM) can be enhanced by an order of magnitude to sub-ppm levels. After mass accuracy enhancement all 211 reference masses have mass errors within 0.5 ppm; only 14 of these are outside the 0.2 ppm error margin. Further demonstration of mass accuracy enhancement is shown on a pre-concentrated urine sample in which evidence for 89 (342 ions) potential hydroxylated and glucuronated DHEA-metabolites is found. Although most DHEA metabolites have low-intensity mass signals, only 11 out of 342 are outside the ±1 ppm error envelop; 272 mass signals have errors below 0.5 ppm (142 below 0.2 ppm). The methodology consists of: (a) a multiple internal lock cor</description><dates><release>2011-01-01T00:00:00Z</release><publication>2011 Mar</publication><modification>2025-05-29T20:56:27.184Z</modification><creation>2025-05-29T20:56:27.184Z</creation></dates><accession>S-EPMC3040356</accession><cross_references><pubmed>21461040</pubmed><doi>10.1007/s11306-010-0230-y</doi></cross_references></HashMap>