<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>85(6)</volume><submitter>Arduino A</submitter><funding>European Association of National Metrology Institutes</funding><pubmed_abstract>&lt;h4>Purpose&lt;/h4>To investigate how the simultaneous exposure to gradient and RF fields affects the temperature rise in patients with a metallic hip prosthesis during an MRI session.&lt;h4>Methods&lt;/h4>In silico analysis was performed with an anatomically realistic human model with CoCrMo hip implant in 12 imaging positions. The analysis was performed at 1.5 T and 3 T, considering four clinical sequences: turbo spin-echo, EPI, gradient-echo, and true fast imaging sequence with steady precession. The exposure to gradient and RF fields was evaluated separately and superposed, by adopting an ad hoc computational algorithm. Temperature increase within the body, rather than specific absorption rate, was used as a safety metric.&lt;h4>Results&lt;/h4>With the exception of gradient-echo, all investigated seq</pubmed_abstract><journal>Magnetic resonance in medicine</journal><pagination>3447-3462</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC7986841</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>Heating of hip joint implants in MRI: The combined effect of RF and switched-gradient fields.</pubmed_title><pmcid>PMC7986841</pmcid><pubmed_authors>Zilberti L</pubmed_authors><pubmed_authors>Chiampi M</pubmed_authors><pubmed_authors>Bottauscio O</pubmed_authors><pubmed_authors>Arduino A</pubmed_authors><pubmed_authors>Hand J</pubmed_authors><pubmed_authors>Bruhl R</pubmed_authors><pubmed_authors>Zanovello U</pubmed_authors></additional><is_claimable>false</is_claimable><name>Heating of hip joint implants in MRI: The combined effect of RF and switched-gradient fields.</name><description>&lt;h4>Purpose&lt;/h4>To investigate how the simultaneous exposure to gradient and RF fields affects the temperature rise in patients with a metallic hip prosthesis during an MRI session.&lt;h4>Methods&lt;/h4>In silico analysis was performed with an anatomically realistic human model with CoCrMo hip implant in 12 imaging positions. The analysis was performed at 1.5 T and 3 T, considering four clinical sequences: turbo spin-echo, EPI, gradient-echo, and true fast imaging sequence with steady precession. The exposure to gradient and RF fields was evaluated separately and superposed, by adopting an ad hoc computational algorithm. Temperature increase within the body, rather than specific absorption rate, was used as a safety metric.&lt;h4>Results&lt;/h4>With the exception of gradient-echo, all investigated seq</description><dates><release>2021-01-01T00:00:00Z</release><publication>2021 Jun</publication><modification>2025-04-04T14:35:19.792Z</modification><creation>2025-04-04T14:35:19.792Z</creation></dates><accession>S-EPMC7986841</accession><cross_references><pubmed>33483979</pubmed><doi>10.1002/mrm.28666</doi></cross_references></HashMap>