<HashMap><database>biostudies-literature</database><scores/><additional><submitter>Teruyama K</submitter><funding>Japan Agency for Medical Research and Development</funding><pagination>77-85</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC8766281</full_dataset_link><repository>biostudies-literature</repository><omics_type>Unknown</omics_type><volume>36(1)</volume><pubmed_abstract>Determination of plasma aldosterone concentrations (PAC) and plasma active renin concentrations (ARC) is essential for the diagnosis of primary aldosteronism (PA). In Japan, although PAC and ARC are measured by radioimmunoassay and immunoradiometric assay, respectively, non-radioisotopic methods with better detection sensitivity, measurement accuracy, and technical simplicity are needed. We developed two-site sandwich chemiluminescent enzyme immunoassays (CLEIAs) to measure both PAC and ARC using monoclonal antibodies immobilized onto ferrite particles. The results of both assays are obtained simultaneously from a single plasma sample within 30 min using a fully automated system. The novel CLEIAs were validated using plasma samples from patients with PA (n = 52) and essential hypertension </pubmed_abstract><journal>Journal of human hypertension</journal><pubmed_title>Novel chemiluminescent immunoassay to measure plasma aldosterone and plasma active renin concentrations for the diagnosis of primary aldosteronism.</pubmed_title><pmcid>PMC8766281</pmcid><funding_grant_id>JP17ek0109122</funding_grant_id><funding_grant_id>JP20ek0109352</funding_grant_id><funding_grant_id>JP19ek0109352</funding_grant_id><pubmed_authors>Naruse M</pubmed_authors><pubmed_authors>Teruyama K</pubmed_authors><pubmed_authors>Kobayashi H</pubmed_authors><pubmed_authors>Tsuiki M</pubmed_authors></additional><is_claimable>false</is_claimable><name>Novel chemiluminescent immunoassay to measure plasma aldosterone and plasma active renin concentrations for the diagnosis of primary aldosteronism.</name><description>Determination of plasma aldosterone concentrations (PAC) and plasma active renin concentrations (ARC) is essential for the diagnosis of primary aldosteronism (PA). In Japan, although PAC and ARC are measured by radioimmunoassay and immunoradiometric assay, respectively, non-radioisotopic methods with better detection sensitivity, measurement accuracy, and technical simplicity are needed. We developed two-site sandwich chemiluminescent enzyme immunoassays (CLEIAs) to measure both PAC and ARC using monoclonal antibodies immobilized onto ferrite particles. The results of both assays are obtained simultaneously from a single plasma sample within 30 min using a fully automated system. The novel CLEIAs were validated using plasma samples from patients with PA (n = 52) and essential hypertension </description><dates><release>2022-01-01T00:00:00Z</release><publication>2022 Jan</publication><modification>2026-05-08T10:35:53.915Z</modification><creation>2025-02-19T01:47:16.291Z</creation></dates><accession>S-EPMC8766281</accession><cross_references><pubmed>33564064</pubmed><doi>10.1038/s41371-020-00465-5</doi></cross_references></HashMap>