{"database":"biostudies-literature","file_versions":[],"scores":null,"additional":{"omics_type":["Unknown"],"volume":["10"],"submitter":["Kuypers KLAM"],"pubmed_abstract":["<h4>Objective</h4>The resistance created by the PEEP-valve of a T-piece resuscitator is bias gas flow dependent and might affect breathing in preterm infants. In this study we investigated the effect of a higher bias gas flow on the imposed inspiratory and expiratory T-piece resistance and expiratory breaking manoeuvres (EBM) in preterm infants during spontaneous breathing on CPAP at birth.<h4>Methods</h4>In a retrospective pre-post implementation study of preterm infants <32 weeks gestation, who were stabilised with a T-piece resuscitator, a bias gas flow of 12 L/min was compared to 8 L/min. All spontaneous breaths on CPAP within the first 10 min of starting respiratory support were analysed on a breath-by-breath basis to determine the breathing pattern of each breath and to calculate the"],"journal":["Frontiers in pediatrics"],"pagination":["817010"],"full_dataset_link":["https://www.ebi.ac.uk/biostudies/studies/S-EPMC8968754"],"repository":["biostudies-literature"],"pubmed_title":["The Effect of a Higher Bias Gas Flow on Imposed T-Piece Resistance and Breathing in Preterm Infants at Birth."],"pmcid":["PMC8968754"],"pubmed_authors":["Kuypers KLAM","Willemsen LA","Kashyap AJ","Cramer SJE","Drevhammar T","Hooper SB","Te Pas AB"],"additional_accession":[]},"is_claimable":false,"name":"The Effect of a Higher Bias Gas Flow on Imposed T-Piece Resistance and Breathing in Preterm Infants at Birth.","description":"<h4>Objective</h4>The resistance created by the PEEP-valve of a T-piece resuscitator is bias gas flow dependent and might affect breathing in preterm infants. In this study we investigated the effect of a higher bias gas flow on the imposed inspiratory and expiratory T-piece resistance and expiratory breaking manoeuvres (EBM) in preterm infants during spontaneous breathing on CPAP at birth.<h4>Methods</h4>In a retrospective pre-post implementation study of preterm infants <32 weeks gestation, who were stabilised with a T-piece resuscitator, a bias gas flow of 12 L/min was compared to 8 L/min. All spontaneous breaths on CPAP within the first 10 min of starting respiratory support were analysed on a breath-by-breath basis to determine the breathing pattern of each breath and to calculate the","dates":{"release":"2022-01-01T00:00:00Z","publication":"2022","modification":"2025-04-04T19:10:19.011Z","creation":"2025-04-04T19:10:19.011Z"},"accession":"S-EPMC8968754","cross_references":{"pubmed":["35372143"],"doi":["10.3389/fped.2022.817010"]}}