<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>10</volume><submitter>Kuypers KLAM</submitter><pubmed_abstract>&lt;h4>Objective&lt;/h4>The resistance created by the PEEP-valve of a T-piece resuscitator is bias gas flow dependent and might affect breathing in preterm infants. In this study we investigated the effect of a higher bias gas flow on the imposed inspiratory and expiratory T-piece resistance and expiratory breaking manoeuvres (EBM) in preterm infants during spontaneous breathing on CPAP at birth.&lt;h4>Methods&lt;/h4>In a retrospective pre-post implementation study of preterm infants &lt;32 weeks gestation, who were stabilised with a T-piece resuscitator, a bias gas flow of 12 L/min was compared to 8 L/min. All spontaneous breaths on CPAP within the first 10 min of starting respiratory support were analysed on a breath-by-breath basis to determine the breathing pattern of each breath and to calculate the</pubmed_abstract><journal>Frontiers in pediatrics</journal><pagination>817010</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC8968754</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>The Effect of a Higher Bias Gas Flow on Imposed T-Piece Resistance and Breathing in Preterm Infants at Birth.</pubmed_title><pmcid>PMC8968754</pmcid><pubmed_authors>Kuypers KLAM</pubmed_authors><pubmed_authors>Willemsen LA</pubmed_authors><pubmed_authors>Kashyap AJ</pubmed_authors><pubmed_authors>Cramer SJE</pubmed_authors><pubmed_authors>Drevhammar T</pubmed_authors><pubmed_authors>Hooper SB</pubmed_authors><pubmed_authors>Te Pas AB</pubmed_authors></additional><is_claimable>false</is_claimable><name>The Effect of a Higher Bias Gas Flow on Imposed T-Piece Resistance and Breathing in Preterm Infants at Birth.</name><description>&lt;h4>Objective&lt;/h4>The resistance created by the PEEP-valve of a T-piece resuscitator is bias gas flow dependent and might affect breathing in preterm infants. In this study we investigated the effect of a higher bias gas flow on the imposed inspiratory and expiratory T-piece resistance and expiratory breaking manoeuvres (EBM) in preterm infants during spontaneous breathing on CPAP at birth.&lt;h4>Methods&lt;/h4>In a retrospective pre-post implementation study of preterm infants &lt;32 weeks gestation, who were stabilised with a T-piece resuscitator, a bias gas flow of 12 L/min was compared to 8 L/min. All spontaneous breaths on CPAP within the first 10 min of starting respiratory support were analysed on a breath-by-breath basis to determine the breathing pattern of each breath and to calculate the</description><dates><release>2022-01-01T00:00:00Z</release><publication>2022</publication><modification>2025-04-04T19:10:19.011Z</modification><creation>2025-04-04T19:10:19.011Z</creation></dates><accession>S-EPMC8968754</accession><cross_references><pubmed>35372143</pubmed><doi>10.3389/fped.2022.817010</doi></cross_references></HashMap>