{"database":"biostudies-literature","file_versions":[],"scores":null,"additional":{"omics_type":["Unknown"],"volume":["6(1)"],"submitter":["Klarenberg H"],"pubmed_abstract":["<h4>Background</h4>A new 72-channel receive array coil and sensitivity encoding, compressed (C-SENSE) and noncompressed (SENSE), were investigated to decrease the number of breath-holds (BHs) for cardiac magnetic resonance (CMR).<h4>Methods</h4>Three-T CMRs were performed using the 72-channel coil with SENSE-2/4/6 and C-SENSE-2/4/6 accelerated short-axis cine two-dimensional balanced steady-state free precession sequences. A 16-channel coil with SENSE-2 served as reference. Ten healthy subjects were included. BH-time was kept under 15 s. Data were compared in terms of image quality, biventricular function, number of BHs, and scan times.<h4>Results</h4>BHs decreased from 7 with C-SENSE-2 (scan time 70 s, 2 slices/BH) to 3 with C-SENSE-4 (scan time 42 s, 4-5 slices/BH) and 2 with C-SENSE-6 ("],"journal":["European radiology experimental"],"pagination":["54"],"full_dataset_link":["https://www.ebi.ac.uk/biostudies/studies/S-EPMC9622972"],"repository":["biostudies-literature"],"pubmed_title":["A 72-channel receive array coil allows whole-heart cine MRI in two breath holds."],"pmcid":["PMC9622972"],"pubmed_authors":["Boekholdt SM","Gosselink M","Lamb HJ","Leiner T","Froeling M","Bakermans AJ","Klarenberg H","Nederveen AJ","Strijkers GJ","Coolen BF"],"additional_accession":[]},"is_claimable":false,"name":"A 72-channel receive array coil allows whole-heart cine MRI in two breath holds.","description":"<h4>Background</h4>A new 72-channel receive array coil and sensitivity encoding, compressed (C-SENSE) and noncompressed (SENSE), were investigated to decrease the number of breath-holds (BHs) for cardiac magnetic resonance (CMR).<h4>Methods</h4>Three-T CMRs were performed using the 72-channel coil with SENSE-2/4/6 and C-SENSE-2/4/6 accelerated short-axis cine two-dimensional balanced steady-state free precession sequences. A 16-channel coil with SENSE-2 served as reference. Ten healthy subjects were included. BH-time was kept under 15 s. Data were compared in terms of image quality, biventricular function, number of BHs, and scan times.<h4>Results</h4>BHs decreased from 7 with C-SENSE-2 (scan time 70 s, 2 slices/BH) to 3 with C-SENSE-4 (scan time 42 s, 4-5 slices/BH) and 2 with C-SENSE-6 (","dates":{"release":"2022-01-01T00:00:00Z","publication":"2022 Nov","modification":"2025-04-19T02:35:04.211Z","creation":"2025-02-19T01:55:52.068Z"},"accession":"S-EPMC9622972","cross_references":{"pubmed":["36316525"],"doi":["10.1186/s41747-022-00305-w"]}}