<HashMap><database>biostudies-literature</database><scores/><additional><omics_type>Unknown</omics_type><volume>6(1)</volume><submitter>Klarenberg H</submitter><pubmed_abstract>&lt;h4>Background&lt;/h4>A new 72-channel receive array coil and sensitivity encoding, compressed (C-SENSE) and noncompressed (SENSE), were investigated to decrease the number of breath-holds (BHs) for cardiac magnetic resonance (CMR).&lt;h4>Methods&lt;/h4>Three-T CMRs were performed using the 72-channel coil with SENSE-2/4/6 and C-SENSE-2/4/6 accelerated short-axis cine two-dimensional balanced steady-state free precession sequences. A 16-channel coil with SENSE-2 served as reference. Ten healthy subjects were included. BH-time was kept under 15 s. Data were compared in terms of image quality, biventricular function, number of BHs, and scan times.&lt;h4>Results&lt;/h4>BHs decreased from 7 with C-SENSE-2 (scan time 70 s, 2 slices/BH) to 3 with C-SENSE-4 (scan time 42 s, 4-5 slices/BH) and 2 with C-SENSE-6 (</pubmed_abstract><journal>European radiology experimental</journal><pagination>54</pagination><full_dataset_link>https://www.ebi.ac.uk/biostudies/studies/S-EPMC9622972</full_dataset_link><repository>biostudies-literature</repository><pubmed_title>A 72-channel receive array coil allows whole-heart cine MRI in two breath holds.</pubmed_title><pmcid>PMC9622972</pmcid><pubmed_authors>Boekholdt SM</pubmed_authors><pubmed_authors>Gosselink M</pubmed_authors><pubmed_authors>Lamb HJ</pubmed_authors><pubmed_authors>Leiner T</pubmed_authors><pubmed_authors>Froeling M</pubmed_authors><pubmed_authors>Bakermans AJ</pubmed_authors><pubmed_authors>Klarenberg H</pubmed_authors><pubmed_authors>Nederveen AJ</pubmed_authors><pubmed_authors>Strijkers GJ</pubmed_authors><pubmed_authors>Coolen BF</pubmed_authors></additional><is_claimable>false</is_claimable><name>A 72-channel receive array coil allows whole-heart cine MRI in two breath holds.</name><description>&lt;h4>Background&lt;/h4>A new 72-channel receive array coil and sensitivity encoding, compressed (C-SENSE) and noncompressed (SENSE), were investigated to decrease the number of breath-holds (BHs) for cardiac magnetic resonance (CMR).&lt;h4>Methods&lt;/h4>Three-T CMRs were performed using the 72-channel coil with SENSE-2/4/6 and C-SENSE-2/4/6 accelerated short-axis cine two-dimensional balanced steady-state free precession sequences. A 16-channel coil with SENSE-2 served as reference. Ten healthy subjects were included. BH-time was kept under 15 s. Data were compared in terms of image quality, biventricular function, number of BHs, and scan times.&lt;h4>Results&lt;/h4>BHs decreased from 7 with C-SENSE-2 (scan time 70 s, 2 slices/BH) to 3 with C-SENSE-4 (scan time 42 s, 4-5 slices/BH) and 2 with C-SENSE-6 (</description><dates><release>2022-01-01T00:00:00Z</release><publication>2022 Nov</publication><modification>2025-04-19T02:35:04.211Z</modification><creation>2025-02-19T01:55:52.068Z</creation></dates><accession>S-EPMC9622972</accession><cross_references><pubmed>36316525</pubmed><doi>10.1186/s41747-022-00305-w</doi></cross_references></HashMap>