Utility and limitations of coherent mapping algorithm utilizing vectors and global propagation patterns in atrial tachycardia.


ABSTRACT:

Background

A novel mapping algorithm utilizing vectors and global patterns of propagation (Coherent™, Biosense Webster) has been developed to help identify the mechanism of atrial tachycardia (AT). We aimed to determine the diagnostic accuracy of coherent mapping compared with that of ripple mapping.

Methods and results

This study included 41 consecutive patients with 84 ATs (47 reentrant and 37 focal ATs). Two independent electrophysiologists confirmed the diagnoses using coherent mapping before the ripple map-guided ablation. AT termination was achieved in 75 of 84 ATs (89%) at first ablation lesion set. Four of the remaining nine ATs, which were terminated before an index radiofrequency (RF) application, were non-inducible after RF delivery at the first lesion set, wherea