Fluid mechanical performance of ureteral stents: The role of side hole and lumen size.
Ontology highlight
ABSTRACT: Ureteral stents are indispensable devices in urological practice to maintain and reinstate the drainage of urine in the upper urinary tract. Most ureteral stents feature openings in the stent wall, referred to as side holes (SHs), which are designed to facilitate urine flux in and out of the stent lumen. However, systematic discussions on the role of SH and stent lumen size in regulating flux and shear stress levels are still lacking. In this study, we leveraged both experimental and numerical methods, using microscopic-Particle Image Velocimetry and Computational Fluid Dynamic models, respectively, to explore the influence of varying SH and lumen diameters. Our results showed that by reducing the SH diameter from 1.1 to 0.4 mm the median wall shear stress levels of the SHs near the ure
SUBMITTER: Zheng S
PROVIDER: S-EPMC10013766 | biostudies-literature | 2023 Mar
REPOSITORIES: biostudies-literature