ABSTRACT: Introduction
The wearable cardioverter defibrillator (WCD) is used to protect patients at risk for sudden cardiac arrest. We examined defibrillation efficacy and safety of a biphasic truncated exponential waveform designed for use in a contemporary WCD in three animal studies and a human study.Methods
Animal (swine) studies: #1: Efficacy comparison of a 170J BTE waveform (SHOCK A) to a 150J BTE waveform (SHOCK B) that approximates another commercially available waveform. Primary endpoint first shock success rate. #2: Efficacy comparison of the two waveforms at attenuated charge voltages in swine at three prespecified impedances. Primary endpoint first shock success rate. #3: Safety comparison of SHOCK A and SHOCK B in swine. Primary endpoint cardiac biomarker level changes