Reproducibility of T and T2 quantification in a multi-vendor multi-site study.


ABSTRACT:

Objective

To evaluate the multi-vendor multi-site reproducibility of two-dimensional (2D) multi-echo spin-echo (MESE) T2 mapping (product sequences); and to evaluate the longitudinal reproducibility of three-dimensional (3D) magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo snapshots (MAPSS) T and T2 mapping (research sequences), and 2D MESE T2 mapping, separated by 6 months, in a multi-vendor multi-site setting.

Methods

Phantoms and volunteers (n = 5 from each site, n = 20 in total) were scanned on four 3 T magnetic resonance (MR) systems from four sites and three vendors (Siemens, General Electric, and Phillips). Two traveling volunteers (3 knees) scanned at all 4 sites at baseline and 6-month follow-up. Data was transferred to one site for centralized processing. Coefficients of variation (CVs) were calculated to evaluate reproducibility.

Results

For baseline 2D MESE T2 measures, average CV were 0.37-2.45% (intra-site) and 5.96% (inter-site) for phantoms, and 3.15-8.49% (intra-site) and 14.16% (inter-site) for volunteers. For longitudinal phantom data, intra-site CVs were 1.42-3.48% for 3D MAPSS T, 1.77-3.56% for 3D MAPSS T2, and 1.02-2.54% for 2D MESE T2. For the longitudinal volunteer data, the intra-site CVs were 2.60-4.86% for 3D MAPSS T, 3.33-7.25% for 3D MAPSS T2, and 3.11-8.77% for 2D MESE T2.

Conclusion

This study demonstrated excellent intra-site reproducibility of 2D MESE T2 imaging, while its inter-site variation was slightly higher than 3D MAPSS T2 imaging (10.06% as previously reported). This study also showed excellent reproducibility of longitudinal T and T2 cartilage quantification, in a multi-vendor multi-site setting for both product 2D MESE T2 and 3D MAPSS T1p/T2 research sequences.

SUBMITTER: Lartey R 

PROVIDER: S-EPMC10016129 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reproducibility of T<sub>1ρ</sub> and T<sub>2</sub> quantification in a multi-vendor multi-site study.

Lartey R R   Nanavati A A   Kim J J   Li M M   Xu K K   Nakamura K K   Shin W W   Winalski C S CS   Obuchowski N N   Bahroos E E   Link T M TM   Hardy P A PA   Peng Q Q   Kim J J   Liu K K   Fung M M   Wu C C   Li X X  

Osteoarthritis and cartilage 20221110 2


<h4>Objective</h4>To evaluate the multi-vendor multi-site reproducibility of two-dimensional (2D) multi-echo spin-echo (MESE) T<sub>2</sub> mapping (product sequences); and to evaluate the longitudinal reproducibility of three-dimensional (3D) magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo snapshots (MAPSS) T<sub>1ρ</sub> and T<sub>2</sub> mapping (research sequences), and 2D MESE T<sub>2</sub> mapping, separated by 6 months, in a multi-vendor multi-site setting  ...[more]

Similar Datasets

| S-EPMC7480266 | biostudies-literature
| S-EPMC9149651 | biostudies-literature
| S-EPMC11651994 | biostudies-literature
| S-EPMC6933042 | biostudies-literature
| S-EPMC7116151 | biostudies-literature
| S-EPMC9082875 | biostudies-literature
| S-EPMC6521472 | biostudies-literature
| S-EPMC6825401 | biostudies-literature
| S-EPMC10365099 | biostudies-literature
| S-EPMC8096323 | biostudies-literature