Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients.


ABSTRACT:

Background

EQ.PET is a software package that overcomes the reconstruction-dependent variation of standard uptake values (SUV). In this study, we validated the use of EQ.PET for harmonizing SUVs between different positron emission tomography/computed tomography (PET/CT) systems and reconstruction algorithms.

Methods

In this retrospective study, 49 patients with various cancers were scanned on a Biograph mCT (mCT) or Gemini TF 16 (Gemini) after [18F]FDG injections. Three groups of patient data were collected: Group 1, patients scanned on mCT or Gemini with data reconstructed using two parameters; Group 2, patients scanned twice on different PET scanners (interval between two scans, 68.9 ± 41.4 days); and Group 3, patients scanned twice using mCT with data reconstructed using different algorithms (interval between two scans, 109.5 ± 60.6 days). The SUVs of the lesions and background were measured, and the tumor-to-background ratios (TBRs) were calculated. In addition, the consistency between the two reconstruction algorithms and confounding factors were evaluated.

Results

In Group 1, the consistency of SUV and TBR between different reconstruction algorithms improved when the EQ.PET filter was applied. In Group 2, by comparing ΔSUV, ΔSUV%, ΔTBR, and ΔTBR% with and without the EQ.PET, the results showed significant differences (P < 0.05). In Group 3, Bland-Altman analysis of ΔSUV with EQ.PET showed an improved consistency relative to that without EQ.PET.

Conclusions

EQ.PET is an efficient tool to harmonize SUVs and TBRs across different reconstruction algorithms. Patients could benefit from the harmonized SUV, ΔSUV, and ΔSUV% for therapy responses and follow-up evaluations.

SUBMITTER: Song Y 

PROVIDER: S-EPMC10017904 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients.

Song Yufei Y   Meng Xiangxi X   Cao Zhen Z   Zhao Wei W   Zhang Yan Y   Guo Rui R   Zhou Xin X   Yang Zhi Z   Li Nan N  

EJNMMI physics 20230315 1


<h4>Background</h4>EQ.PET is a software package that overcomes the reconstruction-dependent variation of standard uptake values (SUV). In this study, we validated the use of EQ.PET for harmonizing SUVs between different positron emission tomography/computed tomography (PET/CT) systems and reconstruction algorithms.<h4>Methods</h4>In this retrospective study, 49 patients with various cancers were scanned on a Biograph mCT (mCT) or Gemini TF 16 (Gemini) after [<sup>18</sup>F]FDG injections. Three  ...[more]

Similar Datasets

| S-EPMC10963814 | biostudies-literature
| S-EPMC10328817 | biostudies-literature
| S-EPMC6034544 | biostudies-literature
| S-EPMC7761511 | biostudies-literature
| S-EPMC11009581 | biostudies-literature
| S-EPMC4647251 | biostudies-literature
| S-EPMC7169460 | biostudies-literature
2011-11-01 | GSE21217 | GEO
2015-02-09 | E-MTAB-3175 | biostudies-arrayexpress
| 2066802 | ecrin-mdr-crc