Performance evaluation of the Asante Rapid Recency Assay for verification of HIV diagnosis and detection of recent HIV-1 infections: Implications for epidemic control.
Ontology highlight
ABSTRACT: We previously described development of a rapid test for recent infection (RTRI) that can diagnose HIV infection and detect HIV-1 recent infections in a single device. This technology was transferred to a commercial partner as Asante Rapid Recency Assay (ARRA). We evaluated performance of the ARRA kits in the laboratory using a well-characterized panel of specimens. The plasma specimen panel (N = 1500) included HIV-1 (N = 570), HIV-2 (N = 10), and HIV-negatives (N = 920) representing multiple subtypes and geographic locations. Reference diagnostic data were generated using the Bio-Rad HIV-1-2-O EIA/Western blot algorithm with further serotyping performed using the Multispot HIV-1/2 assay. The LAg-Avidity EIA was used to generate reference data on recent and long-term infection for HIV-1 pos
SUBMITTER: Yufenyuy EL
PROVIDER: S-EPMC10021762 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature
ACCESS DATA