Unknown

Dataset Information

Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance.


ABSTRACT:

Introduction

A long T2 relaxation time can reflect oedema, and myocardial inflammation when combined with increased plasma troponin levels. Cardiovascular magnetic resonance (CMR) T2 mapping therefore has potential to provide a key diagnostic and prognostic biomarkers. However, T2 varies by scanner, software, and sequence, highlighting the need for standardization and for a quality assurance system for T2 mapping in CMR.

Aim

To fabricate and assess a phantom dedicated to the quality assurance of T2 mapping in CMR.

Method

A T2 mapping phantom was manufactured to contain 9 T1 and T2 (T1|T2) tubes to mimic clinically relevant native and post-contrast T2

SUBMITTER: Topriceanu CC 

PROVIDER: S-EPMC10026458 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature