Unknown

Dataset Information

Paroxysmal Atrial Fibrillation Ablation Using a Novel Variable-Loop Biphasic Pulsed Field Ablation Catheter Integrated With a 3-Dimensional Mapping System: 1-Year Outcomes of the Multicenter inspIRE Study.


ABSTRACT:

Background

The inspIRE study (Study for Treatment of Paroxysmal Atrial Fibrillation [PAF] by Pulsed Field Ablation [PFA] System With Irreversible Electroporation [IRE]) evaluated safety and effectiveness of a fully integrated biphasic pulsed field ablation (PFA) system with a variable-loop circular catheter for the treatment of drug-refractory paroxysmal atrial fibrillation.

Methods

Subjects underwent pulmonary vein (PV) isolation with the PFA system, using at least 12 applications per vein; adenosine/isoproterenol was administered to confirm entrance block. Wave I assessed initial safety, including for esophageal lesions, silent cerebral lesions, and PV stenosis. Wave II (pivotal phase) tested (1) primary safety, incidence of early-onset primary adverse events, and (2) prim

SUBMITTER: Duytschaever M 

PROVIDER: S-EPMC10026968 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature