Delivering Low Tidal Volume With Anesthesia and ICU Ventilators in a Neonatal Lung Model.


ABSTRACT:

Background

Mechanical ventilation of the neonate requires ventilators than can deliver precise and accurate tidal volume (VT) and PEEP to avoid lung injury. Due to small neonatal VT and the disproportionate effect of endotracheal tube leak in these patients, accomplishing precise and accurate VT delivery is difficult. Whereas neonatal ICU ventilators are validated in this population, thorough studies testing the performance of anesthesia ventilators in delivering small VT in neonates are lacking.

Methods

Three anesthesia ventilators, Dräger Apollo, GE Avance, and Getinge Flow-i; and 2 ICU ventilators, Medtronic PB980 and Nihon Kohden NKV-550, were tested under volume control mode at VT of 5, 20, 40, and 60 mL. Three comb