Ontology highlight
ABSTRACT: Purpose
The purpose of this study was to compare total keratometry (TK) and posterior keratometry (PK) obtained by two swept-source optical biometers (IOLMaster 700 and CASIA2) and one Scheimpflug-based topography (Pentacam AXL).Methods
The TK and PK in cataract surgery candidates obtained by IOLMaster 700, CASIA2, and Pentacam AXL were compared. Intraclass correlation coefficients (ICCs), limit of agreement, and Bland-Altman plots were used to assess the agreement.Results
One hundred two patients with a mean age of 68.21 ± 8.70 years were included. There were significant differences among IOLMaster 700, CASIA2, and Pentacam AXL in the mean TK (TKm) (44.23 ± 1.59 diopters [D] vs. 43.25 ± 1.53 D vs. 43.94 ± 1.68 D; all P < 0.001), mean PK (PKm; -5.90 ± 0.24 D vs. -6.
SUBMITTER: Jin A
PROVIDER: S-EPMC10029765 | biostudies-literature | 2023 Mar
REPOSITORIES: biostudies-literature