Development and Application of Desorption Electrospray Ionization Mass Spectrometry for Historical Dye Analysis.
Ontology highlight
ABSTRACT: A desorption electrospray ionization (DESI) source was built and attached to a Bruker 7T SolariX FT-ICR-MS for the in situ analysis of 14 early synthetic dyestuffs. Optimization using silk and wool cloths dyed with rhodamine B concluded that when using a commercial electrospray emitter (part number: 0601815, Bruker Daltonik), a nebulizing gas (N2) pressure of 3.9 bar and a sprayer voltage of 4.5 kV (positive ionization mode) or 4.2 kV (negative ionization mode), a solvent system of 3:1 v/v ACN:H2O, and a sprayer incident angle, α, of 35° gave the highest signal-to-noise ratios on both silk and wool for the samples investigated. The system was applied to modern early synthetic dye references on silk and wool as well as historical samples from the 1893 edition of
SUBMITTER: Sandstrom E
PROVIDER: S-EPMC10034744 | biostudies-literature | 2023 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA