ABSTRACT: Background and purpose
Low-field magnetic resonance imaging (MRI) may offer specific advantages over high-field MRI, e.g. lower susceptibility-dependent distortions and simpler installation. The study aim was to evaluate if a novel 0.55 T MRI scanner provides sufficient image accuracy and quality for radiotherapy (RT) treatment planning.Material and methods
The geometric accuracy of images acquired at a low-field MRI scanner was evaluated in phantom measurements regarding gradient non-linearity-related distortions. Patient-induced B0-susceptibility changes were investigated via B0-field-mapping in ten volunteers. Patients were positioned in RT-setup using a 3D-printed insert for the head/neck-coil that was tested for sufficient signal-to-noise-ratio (SNR). The suitability o