Parametric study of a microdialysis probe and study of depletion effect using ethanol as a test analyte.


ABSTRACT: Although microdialysis is a common in vivo sampling technique, a detailed characterization of the performance of a microdialysis probe used for sampling ethanol molecules has not been conducted. In this work, experimental and computational investigations were carried out to quantitatively study ethanol diffusion characteristics for home-made and commercially available probes. Probe efficiency, i.e. recovery rate (defined as the ethanol concentration in the dialysate to that in the external medium surrounding the probe) was used to characterize the performance. The recovery rate was measured at different perfusion flow rates (0.1, 0.2, 0.5, 1, 1.5, 2 μL/min) and external ethanol concentrations (1, 2.5, 5, 10, 20 mM) with controlled environmental conditions. Effect of temperature was also in

SUBMITTER: Lee TA 

PROVIDER: S-EPMC10039669 | biostudies-literature | 2022 Dec