Freestanding lipid bilayer tensiometer for the study of mechanosensitive ion channels.
Ontology highlight
ABSTRACT: Mechanical forces modify the cell membrane potential by opening mechanosensitive ion channels. We report the design and construction of a lipid bilayer tensiometer to study channels that respond to lateral membrane tension, [Formula: see text] , in the range 0.2 to 1.4 [Formula: see text] (0.8 to 5.7 [Formula: see text] ). The instrument consists of a black-lipid-membrane bilayer, a custom-built microscope, and a high-resolution manometer. Values of [Formula: see text] are obtained from the determination of the bilayer curvature as a function of applied pressure by means of the Young-Laplace equation. We demonstrate that [Formula: see text] can be determined by calculating the bilayer radius of curvature from fluorescence microscopy imaging or from measurements of the bilayer's electrical