Evaluating clinical near-infrared surgical camera systems with a view to optimizing operator and computational signal analysis.


ABSTRACT:

Significance

As clinical evidence on the colorectal application of indocyanine green (ICG) perfusion angiography accrues, there is also interest in computerizing decision support. However, user interpretation and software development may be impacted by system factors affecting the displayed near-infrared (NIR) signal.

Aim

We aim to assess the impact of camera positioning on the displayed NIR signal across different open and laparoscopic camera systems.

Approach

The effects of distance, movement, and target location (center versus periphery) on the displayed fluorescence signal of different systems were measured under electromagnetic stereotactic guidance from an ICG-albumin model and in vivo during surgery.

Results

Systems displayed distinct fluorescence performances with variance apparent with scope optical lens configuration (0 deg versus 30 deg), movement, target positioning, and distance. Laparoscopic system readings fitted inverse square function distance-intensity curves with one device and demonstrated a direction dependent sigmoid curve. Laparoscopic cameras presented central targets as brighter than peripheral ones, and laparoscopes with angled optical lens configurations had a diminished field of view. One handheld open system also showed a distance-intensity relationship, whereas the other maintained a consistent signal despite distance, but both presented peripheral targets brighter than central ones.

Conclusions

Optimal clinical use and signal computational development requires detailed appreciation of system behaviors.

SUBMITTER: Dalli J 

PROVIDER: S-EPMC10050972 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluating clinical near-infrared surgical camera systems with a view to optimizing operator and computational signal analysis.

Dalli Jeffrey J   Jindal Abhinav A   Gallagher Gareth G   Epperlein Jonathan P JP   Hardy Niall P NP   Malallah Ra'ed R   O'Donoghue Kilian K   Cantillon-Murphy Padraig P   Mac Aonghusa Pól G PG   Cahill Ronan A RA  

Journal of biomedical optics 20230329 3


<h4>Significance</h4>As clinical evidence on the colorectal application of indocyanine green (ICG) perfusion angiography accrues, there is also interest in computerizing decision support. However, user interpretation and software development may be impacted by system factors affecting the displayed near-infrared (NIR) signal.<h4>Aim</h4>We aim to assess the impact of camera positioning on the displayed NIR signal across different open and laparoscopic camera systems.<h4>Approach</h4>The effects  ...[more]

Similar Datasets

| S-EPMC4334690 | biostudies-literature
| S-EPMC4358042 | biostudies-other
| S-EPMC4595515 | biostudies-literature
| S-EPMC9899360 | biostudies-literature
| S-EPMC7935778 | biostudies-literature
| S-EPMC9370963 | biostudies-literature
| S-EPMC10232307 | biostudies-literature
| S-EPMC5038671 | biostudies-literature
| S-EPMC5059626 | biostudies-literature
| S-EPMC9116886 | biostudies-literature