BioMimics 3D Stent in Femoropopliteal Lesions: 3-Year Outcomes with Propensity Matching for Drug-Coated Balloons.


ABSTRACT:

Background

Through its helical centreline geometry, the BioMimics 3D vascular stent system is designed for the mobile femoropopliteal region, aiming to improve long-term patency and the risk of stent fractures.

Methods

MIMICS 3D is a prospective, European, multi-centre, observational registry to evaluate the BioMimics 3D stent in a real-world population through 3 years. A propensity-matched comparison was performed to investigate the effect of the additional use of drug-coated balloons (DCB).

Results

The MIMICS 3D registry enrolled 507 patients (518 lesion, length 125.9 ± 91.0 mm). At 3 years, the overall survival was 85.2%, freedom from major amputation 98.5%, freedom from clinically driven target lesion revascularisation 78.0%, and primary patency 70.2%. The propensity-matched cohort included 195 patients in each cohort. At 3-year follow-up, there was no statistically significant difference in clinical outcomes, such as overall survival (87.9% in the DCB vs. 85.1% in the no DCB group), freedom from major amputation (99.4% vs. 97.2%), clinically driven TLR (76.4% vs. 80.3%), and primary patency (68.5% vs. 74.4%).

Conclusion

The MIMICS 3D registry showed good 3-year outcomes of the BioMimics 3D stent in femoropopliteal lesions, demonstrating the safety and performance of this device under real-world conditions, whether used alone or in combination with a DCB.

SUBMITTER: Piorkowski M 

PROVIDER: S-EPMC10051351 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

BioMimics 3D Stent in Femoropopliteal Lesions: 3-Year Outcomes with Propensity Matching for Drug-Coated Balloons.

Piorkowski Michael M   Zeller Thomas T   Rammos Christos C   Deloose Koen K   Hertting Klaus K   Sesselmann Volker V   Tepe Gunnar G   Gaines Peter P   Lichtenberg Michael M  

Journal of cardiovascular development and disease 20230316 3


<h4>Background</h4>Through its helical centreline geometry, the BioMimics 3D vascular stent system is designed for the mobile femoropopliteal region, aiming to improve long-term patency and the risk of stent fractures.<h4>Methods</h4>MIMICS 3D is a prospective, European, multi-centre, observational registry to evaluate the BioMimics 3D stent in a real-world population through 3 years. A propensity-matched comparison was performed to investigate the effect of the additional use of drug-coated bal  ...[more]

Similar Datasets

| S-EPMC10871603 | biostudies-literature
| S-EPMC9743239 | biostudies-literature
| S-EPMC11387717 | biostudies-literature
| S-EPMC10515524 | biostudies-literature
| S-EPMC10838477 | biostudies-literature
| S-EPMC7360381 | biostudies-literature
| S-EPMC9537625 | biostudies-literature
| S-EPMC10630168 | biostudies-literature
| S-EPMC11681599 | biostudies-literature
| S-EPMC8270410 | biostudies-literature