Unknown

Dataset Information

Using Capillary Whole Blood to Quantitatively Measure Ferritin: A Validation Trial of a Point-of-Care System.


ABSTRACT: Iron deficiency is a public health problem with devastating health, developmental and behavioral effects which often exacerbated due to affordability and access to screening and diagnosis. Using IronScan™ a portable, point-of-care diagnostic system capable of quantitatively measuring ferritin in blood, we validated IronScan™ ferritin measurements using whole blood and serum with a lab-based, regulator-approved analytical device for measuring ferritin in venous serum. Capillary (finger stick) and venous whole blood samples were obtained from 44 male and female volunteers. Venous serum (vSer) ferritin concentrations were measured on Immulite 2000 Xpi (gold standard). Capillary whole blood (cWB), venous whole blood (vWB), and vSer ferritin levels were measured by IronScan™

SUBMITTER: Fiddler JL 

PROVIDER: S-EPMC10052100 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature