In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating.


ABSTRACT:

Background and objectives

Fractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clinical outcome but are difficult to obtain from planar X-rays. A novel implant positioning technology for proximal humerus plating (Xin1) outputs screw lengths suggestions and plate position based on hole projections in conventional X-ray images. This study investigated the performance of a prototype Xin1 system in a postmortem (in vitro) experiment as well as in a clinical handling test.

Materials and methods

For in vitro testing, twelve shoulders from six anatomical specimens were