ABSTRACT: Purpose
Robot assistance in membrane peeling may improve precision and dexterity or prevent complications by task automation. To design robotic devices, surgical instruments' velocity, acceptable position/pose error, and load ability need to be precisely quantified.Methods
A fiber Bragg grating and inertial sensors are attached to forceps. Data collected from forceps and microscope images are used to quantify a surgeon's hand motion (tremor, velocity, posture perturbation) and operation force (voluntary and involuntary) in inner limiting membrane peeling. All peeling attempts are performed on rabbit eyes in vivo by expert surgeons.Results
The root mean square (RMS) of the tremor amplitude is 20.14 µm (transverse, X), 23.99 µm (transverse, Y), and 11.68 µm (axial, Z)