Depth profiling and standardization from the back side of a sample for accurate analyses: Emphasis on quantifying low-fluence, shallow implants in diamond-like carbon.


ABSTRACT:

Rationale

Back-side thinning of wafers is used to eliminate issues with transient sputtering when analyzing near-surface element distributions. Precise and accurate calibrated implants are created by including a standard reference material during the implantation. Combining these methods allows accurate analysis of low-fluence, shallow features even if matrix effects are a concern.

Methods

Implanted Na (<2.0 × 1011 ions/cm2 , peaking <50 nm) in diamond-like carbon (DLC) film on silicon (solar wind returned by NASA's Genesis mission) was prepared for measurement as follows. Implanted surfaces of samples were epoxied to wafers and back-side-thinned using physical or chemical methods. Thinned samples were then implanted with reference ions for accurate qua

SUBMITTER: Rieck KD 

PROVIDER: S-EPMC10078471 | biostudies-literature | 2023 Mar