Early detection of oliguric events in critically ill patients in the ICU with a novel continuous urine flow measurement device: results of an initial validation study.


ABSTRACT:

Objectives

Urine output is used to evaluate fluid status and is an important marker for acute kidney injury (AKI). Our primary aim was to validate a new automatic urine output monitoring device by comparison to the current practice - the standard urometer.

Methods

We conducted a prospective observational study in three ICUs. Urine flow measurements by Serenno Medical Automatic urine output measuring device (Serenno Medical, Yokneam, Israel) were compared to standard urometer readings taken automatically at 5-minutes intervals by a camera, and to hourly urometer readings by the nurses, both over 1 to 7 days. Our primary outcome was the difference between urine flow assessed by the Serenno device and reference camera-derived measurements (Camera). Our secondary outcome was the