Comparison of Radifocus versus Silverway Guidewires for Percutaneous Radial Angiography Following Failed Use of a J-Tip Guidewire.


ABSTRACT: BACKGROUND During transradial coronary angiography, when conventional J-tip wires fail to deliver catheters to the aortic root due to anatomical obstacles, additional hydrophilic wires, such as Radifocus (Terumo) or Silverway (Asahi), are used. We recently showed that the Silverway guidewire was effective at delivering the catheter to the aortic root. In this study, we aimed to compare the efficacy and safety of Radifocus and Silverway guidewires in 100 patients after failed use of the J-tip guidewire. MATERIAL AND METHODS After patients had a failure of a conventional J-tip wire to reach the aortic root, 100 patients were 1:1 randomized to either the Silverway or Radifocus wire. All patients with failure of the J-tip wire were eligible. The primary endpoint was the time between wire entry