Bionate® nucleus disc replacement: bench testing comparing two different designs.


ABSTRACT:

Background

Intervertebral disc nucleus degeneration initiates a degenerative cascade and can induce chronic low back pain. Nucleus replacement aims to replace the nucleus while the annulus is still intact. Over time, several designs have been introduced, but the definitive solution continues to be elusive. Therefore, we aimed to create a new nucleus replacement that replicates intact intervertebral disc biomechanics, and thus has the potential for clinical applications.

Materials and methods

Two implants with an outer ring and one (D2) with an additional midline strut were compared. Static and fatigue tests were performed with an INSTRON 8874 following the American Society for Testing and Materials F2267-04, F2346-05, 2077-03, D2990-01, and WK4863. Implant stiffness was anal