Ontology highlight
ABSTRACT: Background and purpose
For pelvic magnetic resonance imaging (MRI)-only radiotherapy the use of receiver coil bridges (CB) is recommended to avoid deformation of the patient. Development in coil technology has enabled lightweight, flexible coils. In this work we evaluate the effects of a lightweight coil in a pelvic MRI-only radiotherapy workflow.Materials and methods
Twenty-one patients, referred to prostate MRI-only radiotherapy, were included. Images were acquired with and without CB. Anatomical deformation from the on-patient coil placement was measured in the anterior-posterior (AP) and left-right (LR) direction. The change in signal-to-noise ratio (SNR) was measured in phantom and in vivo.The clinical treatment plan, created on the image with CB, was transferred and r
SUBMITTER: Scherman J
PROVIDER: S-EPMC10091035 | biostudies-literature | 2023 Apr
REPOSITORIES: biostudies-literature