A cryogenic 14-channel 13 C receiver array for 3T human head imaging.


ABSTRACT:

Purpose

This article presents a novel 14-channel receive-only array for 13 C human head imaging at 3 T that explores the SNR gain by operating at cryogenic temperature cooled by liquid nitrogen.

Methods

Cryostats are developed to evaluate single-coil bench SNR performance and cool the 14-channel array with liquid nitrogen while having enough thermal insulation between the coils and the sample. The temperature distribution for the coil array is measured. Circuits are adapted to the -189°C environment and implemented in the 14-channel array. 13 C images are acquired with the array at cryogenic and room temperature in a 3T scanner.

Results

Compared with room temperature, the array at cryogenic temperature provides 27%-168% SNR improvement over all v