Ultra-high-performance supercritical fluid chromatography-mass spectrometry for the analysis of organic contaminants in sediments.


ABSTRACT: A nontarget screening method was developed based on D-optimal designs for ultra-high performance supercritical fluid chromatography with positive and negative electrospray ionization mode mass spectrometry. A mixture of organic contaminants such as pesticides, steroids, surfactants, phenolic and fatty acids, and polycyclic aromatic hydrocarbon derivatives, was used for the optimization. An aprotic mixture of dichloromethane and acetone [3:1] performed overall best as the injection solvent. The highest peak capacities (n) were accomplished at the shallowest gradient (1%B/min), ammonium formate (n = 378 in negative ionization mode), or ammonium acetate (n = 327 in positive ionization mode) in methanol as the modifier. Capillary voltage, make-up solvent flow rate, water, and additive concentr