A cross-calibration study using a novel dual X-ray absorptiometry system for bone mineral density measurements with the European Spine Phantom.


ABSTRACT:

Background

For bone health assessment, dual-energy X-ray absorptiometry (DEXA) is recommended to measure bone mineral content and areal bone mineral density (aBMD) in the lumbar spine. However, intermachine differences were not taken into account when developing these recommendations. According to the International Society of Clinical Densitometry (ISCD), phantom-based cross-calibration is adequate after replacing the DEXA system from a different manufacturer. For different DEXA equipment, individual calibration equations were found to be necessary to fit the observed values with the given densities.

Methods

The BMD European Spine Phantom (ESP) measurements (L1, L2, and L3) were assessed on 3 machines. We used the Welch test in the one-way analysis of variance (ANOVA) with a