Ontology highlight
ABSTRACT: Purpose
The Zepto nano-pulse precision capsulotomy is a novel device for capsulorhexis formation during cataract surgery. Few complications or challenges have been reported while using this device. The purpose of this paper is to highlight two intra-operative challenges that were encountered while using the Zepto device.Observations
CASE 1 - A 65-year-old with advanced primary open angle glaucoma (POAG) and an in situ Ahmed Glaucoma Valve located in the anterior chamber. During a planned phacoemulsification procedure, the tube became trapped between the suction cup of the Zepto device and the lens, resulting in a sudden complete collapse of the anterior chamber. The procedure was completed after appropriate interventions. On post-operative day 1 Descemet folds were v
SUBMITTER: Kalambe AS
PROVIDER: S-EPMC10106479 | biostudies-literature | 2023 Jun
REPOSITORIES: biostudies-literature