A current limitation of single chamber implantable cardioverter defibrillators (ICDs) is the lack of an atrial lead to reliably detect atrial fibrillation (AF) episodes. A novel ventricular based atrial fibrillation (VBAF) detection algorithm was created for single chamber ICDs to assess R-R variability for detection of AF.Methods
Patients implanted with Visia AF™ ICDs were prospectively enrolled in the Medtronic Product Surveillance Registry from December 15, 2015 to January 23, 2019 and followed with at least 30 days of monitoring with the algorithm. Time to device-detected daily burden of AF ≥ 6 min, ≥6 h, and ≥23 h were reported. Clinical actions after device-detected AF were recorded.Results
A total of 291 patients were enrolled with a mean follow-