Development and Validation of Two Optimized Multiplexed Serologic Assays for the 9-Valent Human Papillomavirus Vaccine Types.
Ontology highlight
ABSTRACT: Two multiplex immunoassays are routinely used to assess antibody responses in clinical trials of the 9-valent human papillomavirus (9vHPV) vaccine. The HPV6/11/16/18/31/33/45/52/58 competitive Luminex immunoassay (HPV-9 cLIA) and HPV6/11/16/18/31/33/45/52/58 total immunoglobulin G Luminex immunoassay are used for measurements of immunogenicity. Following their initial validation in 2010, both assays were redeveloped, and several parameters were optimized, including the coating concentration of virus-like particles, type of Luminex microspheres, serum sample and reference standard diluent, reference standard starting dilution and titration series, and vendor and concentration of the phycoerythrin-labeled antibodies. Validation studies evaluated the assay performance parameters, including th
SUBMITTER: Nolan KM
PROVIDER: S-EPMC10117101 | biostudies-literature | 2023 Apr
REPOSITORIES: biostudies-literature