Assessment of Nit-Occlud atrial septal defect occluder device healing process using micro-computed tomography imaging.
Ontology highlight
ABSTRACT: After percutaneous implantation of a cardiac occluder, a complex healing process leads to the device coverage within several months. An incomplete device coverage increases the risk of device related complications such as thrombosis or endocarditis. We aimed to assess the device coverage process of atrial septal defect (ASD) occluders in a chronic sheep model using micro-computed tomography (micro-CT). After percutaneous creation of an ASD, 8 ewes were implanted with a 16-mm Nit-Occlud ASD-R occluder (PFM medical, Cologne, Germany) and were followed for 1 month (N = 3) and 3 months (N = 5). After heart explant, the device coverage was assessed using micro-CT (resolution of 41.7 μm) and was compared to histological analysis. The micro-CT image reconstruction was performed in 2D and 3D allow
SUBMITTER: Perdreau E
PROVIDER: S-EPMC10124873 | biostudies-literature | 2023
REPOSITORIES: biostudies-literature
ACCESS DATA