ABSTRACT: Background
Healthcare workers treating SARS-CoV-2 patients are at risk of infection by respiratory exposure to patient-emitted, virus-laden aerosols. Source control devices such as ventilated patient isolation hoods have been shown to limit the dissemination of non-infectious airborne particles in laboratory tests, but data on their performance in mitigating the airborne transmission risk of infectious viruses are lacking.Aim
We used an infectious airborne virus to quantify the ability of a ventilated hood to reduce infectious virus exposure in indoor environments.Methods
We nebulized 109 plaque forming units (pfu) of bacteriophage PhiX174 virus into a ∼30-m3 room when the hood was active or inactive. The airborne concentration of infectious vi