Influence of image reconstruction on quantitative cardiac 15O-water positron emission tomography.


ABSTRACT:

Background

The impact on quantitative 15O-water PET/CT of a wide range of different reconstruction settings, including regularized reconstruction by block-sequential regularized expectation maximization (BSREM), was investigated.

Methods

Twenty clinical stress scans from patients referred for assessment of myocardial ischemia were included. Patients underwent a 4-min dynamic stress PET scan with 15O-water on a digital PET/CT scanner. Twenty-two reconstructions were generated from each scan and a clinical reconstruction was used as reference. Varied parameters were number of iterations, filter, exclusion of time-of-flight and point-spread function, and regularization parameter with BSREM. Analyses were performed in aQuant utilizing two different methods and resulting regional myocardial blood flow (MBF), perfusable tissue fraction (PTF), and transmural MBF (MBFt) values were evaluated.

Results

Across the two analyses, correlations toward the reference reconstruction were strong for all parameters (ρ ≥ 0.83). Using automated analysis and the diagnostic threshold of hyperemic MBF at 2.3 mL⋅g-1⋅min-1, diagnosis was unchanged irrespective of reconstruction method in all patients except for one, where only four of the most extreme reconstruction methods resulted in a change of diagnosis.

Conclusion

The low sensitivity of MBF values to reconstruction method and, as previously shown, scanner type and PET/CT misalignment, confirms that diagnostic hyperemic MBF cutoff values can be consistently used for 15O-water.

SUBMITTER: Nordstrom J 

PROVIDER: S-EPMC10126040 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Influence of image reconstruction on quantitative cardiac <sup>15</sup>O-water positron emission tomography.

Nordström Jonny J   Lindström Elin E   Kero Tanja T   Sörensen Jens J   Lubberink Mark M  

Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 20220804 2


<h4>Background</h4>The impact on quantitative <sup>15</sup>O-water PET/CT of a wide range of different reconstruction settings, including regularized reconstruction by block-sequential regularized expectation maximization (BSREM), was investigated.<h4>Methods</h4>Twenty clinical stress scans from patients referred for assessment of myocardial ischemia were included. Patients underwent a 4-min dynamic stress PET scan with <sup>15</sup>O-water on a digital PET/CT scanner. Twenty-two reconstruction  ...[more]

Similar Datasets

| S-EPMC9345798 | biostudies-literature
| S-EPMC5306314 | biostudies-literature
| S-EPMC6037289 | biostudies-literature
| S-EPMC11425657 | biostudies-literature
| S-EPMC7161423 | biostudies-literature
| S-EPMC10036920 | biostudies-literature
2011-11-01 | GSE21217 | GEO
2015-02-09 | E-MTAB-3175 | biostudies-arrayexpress
| S-EPMC7273977 | biostudies-literature
| S-EPMC9380568 | biostudies-literature