Accuracy of hemodynamic parameters derived by GE E-PiCCO in comparison with PiCCO® in patients admitted to the intensive care unit.


ABSTRACT: To evaluate the agreement and accuracy of a novel advanced hemodynamic monitoring (AHM) device, the GE E-PiCCO module, with the well-established PiCCO® device in intensive care patients using pulse contour analysis (PCA) and transpulmonary thermodilution (TPTD). A total of 108 measurements were performed in 15 patients with AHM. Each of the 27 measurement sequences (one to four per patient) consisted of a femoral and a jugular indicator injection via central venous catheters (CVC) and measurement using both PiCCO (PiCCO® Jug and Fem) and GE E-PiCCO (GE E-PiCCO Jug and Fem) devices. For statistical analysis, Bland-Altman plots were used to compare the estimated values derived from both devices. The cardiac index measured via PCA (CIpc) and TPTD (CItd) was the only parameter that fulfilled a