Unknown

Dataset Information

New Trend in Toxicological Screening Using Volumetric Absorptive Microsampling (VAMS) and High-Resolution Mass Spectrometry (HR/MS) Combination.


ABSTRACT: In toxicology, screenings are routinely performed using chromatographic methods coupled to detection systems such as high-resolution mass spectrometry (HR/MS). The increase in specificity and sensitivity of HRMS is responsible for the development of methods for alternative samples such as Volumetric Adsorptive Micro-Sampling. Whole blood overloaded with 90 drugs was sampled with 20 µL MitraTM to optimize the pre-analytical step as well as to determine the identification limits of drugs. Elution of chemicals was carried out in a solvent mixture through agitation and sonication. After dissolution, 10 μL was injected into the chromatographic system coupled to the OrbitrapTM HR/MS. Compounds were confirmed against the laboratory library. The clinical feasibility was asses

SUBMITTER: Houze P 

PROVIDER: S-EPMC10141006 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature