Detecting True Change in Keratoconus after Intracorneal Ring Segment Implantation.


ABSTRACT: Confirming the progression of keratoconus is of paramount relevance to providing the appropriate treatment. Real change should be considered consistent over time. It must be greater than the variability of the measurement of the device used to monitor the cornea. The present study aimed to assess the intraobserver repeatability and intersession reproducibility of a Scheimpflug camera in measuring corneal parameters in virgin keratoconus and intrastromal corneal ring segments (ICRS) implantation eyes to discriminate real change from measurement noise. Sixty keratoconus and 30 ICRS eyes were included. Corneal parameters were determined in three consecutive measurements and were repeated 2 weeks later. The precision within the same session for all parameters was better in the keratoconic eyes