Determination of the erythrocyte sedimentation rate using the hematocrit-corrected aggregation index and mean corpuscular volume.


ABSTRACT:

Background

Determination of the erythrocyte sedimentation rate (ESR) by measurement of erythrocyte aggregation is an alternative to the Westergren method and can be performed rapidly. However, its principle is opaque and the ESR values obtained can deviate from Westergren method values (WG ESR) due to hematocrit. Furthermore, WG ESR is affected by particle size, but no studies have examined the effect of individual mean corpuscular volumes (MCVs).

Methods

Simultaneous measurement of the erythrocyte aggregation index (AI) over a 5-s interval and determination of the complete blood count in 80 μL blood from 203 patients were performed (hematocrit, 21.4%-52.3%; MCV, 62.7-114.1 fL). ESR values were calculated with the hematocrit-corrected AI (HAI) for comparison with WG ESR. We